Cleaning and disinfecting
Table 1: Approved cleaners
Product
b
Sani-Cloth AF3
Sani-Cloth Plus
PI-Spray II
a
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum period
of time.
b
Qualified for use as an intermediate-level disinfectant for mycobacteria.
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
Table 2:
High-level compatible disinfectants
Disinfectant
Compatible transducers
Cidex
C11x, HFL38xi, HFL50x, HSL25x,
ICTx, L25x, L38xi, P10x, P11x,
rC60xi, rP19x
Cidex OPA
C11x, HFL38xi, HFL50x, HSL25x,
ICTx, L25x, L38xi, L52x, P10x,
rC60xi, rP19x
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
4
Compatible Transducer
C11x, HFL38xi, HFL50x, HSL25x,
ICTx, L25x, L38xi, P10x, P11x,
rC60xi, rP19x
HFL50x, L52x
C11x, HFL38xi, HFL50x, HSL25x,
ICTx, L25x, L38xi, L52x, P10x,
P11x, rC60xi, rP19x
Minimum wet contact time
3 minutes
3 minutes
10 minutes
www.sonosite.com
for a complete list of the most
Temperature
Disinfectant Soak Duration
o
o
45 minutes
25
C, 77
F
o
o
12 minutes
20
C, 68
F
www.sonosite.com
for a complete list of the most
a
Cleaning and disinfecting