Battery Indication - arjo Flowtron ACS900 Instrucciones De Uso

Ocultar thumbs Ver también para Flowtron ACS900:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Mains Power
Indication

Battery Indication

Wait/Initialising
Therapy Time
The pump is connected to the mains power
supply.
The pump is NOT connected to the mains
power supply.
When the pump is connected to the mains power
supply:
• If the battery is fully charged, the indicator will be
static and show "full".
• If the battery is not fully charged, the indicator will
scroll from "Battery Empty" to the current battery
capacity for example "Battery ½ Full", to indicate
that the pump is being charged.
When the pump is NOT connected to the mains power
supply, the indicator is static and shows the
approximate charge remaining in the battery.
Battery Empty
Battery ½ Full
Battery Full
NOTE
If the pump is not connected to the mains power
supply and the battery is fully charged and in good
condition, the pump will continue to operate normally
for approximately 12 hours for calf, and calf & thigh
garments, and for approximately 6 hours for foot
garments.
NOTE
The battery is not user replaceable and must be
replaced as part of the service procedure once end of
life.
This is a rotating 6-segment circular icon, which
indicates that the pump is initialising after power up.
This shows the accumulated therapy time in hours and
minutes. Refer to "Resetting the Therapy Time" on
page 20 to clear the therapy time.
NOTE
This is the total time of delivered therapy since the
therapy time was last reset.
11
Battery ¼ Full
Battery ¾ Full

Publicidad

Tabla de contenido
loading

Tabla de contenido