Condition
Battery Fault
Description and Corrective Action
The Battery Fault alarm is activated if the pump
detects a fault in the internal battery.
If the pump is in Standby, on battery power:
• An alarm is activated which can be muted.
• The battery icon flashes and shows the last
known charge remaining in the battery.
• The "spanner" symbol flashes.
• Therapy cannot be started.
• Call the service engineer.
If the pump is in the Run state, on battery power:
• An alarm is activated which can be muted.
• The battery icon flashes and shows the last
known charge remaining in the battery.
• The "spanner" symbol flashes.
• Therapy can continue until the battery has
expired.
• Call the service engineer.
If the pump is in Standby, on mains power:
• An alarm is activated which can be muted.
• The battery icon flashes and shows the last
known charge remaining in the battery.
• The "spanner" symbol flashes.
• Therapy can be started.
• Call the service engineer.
If the pump is in the Run state, on mains power:
• An alarm is activated which can be muted.
• The battery icon flashes and shows the last
known charge remaining in the battery.
• The "spanner" symbol flashes.
• Therapy can continue.
• Call the service engineer.