Condition
Garment Removed from the Limb
Service Required
Description and Corrective Action
This warning is activated after a few minutes if the
garment is taken off the patient's leg while the pump
is in run mode. It shows that garment 1 has been
taken off the leg. The leg icon flashes and the
duration that the garment has been removed from
the limb is displayed in minutes in the bottom right
corner.
Additionally the green LED indicators will flash. The
therapy time will be paused, and the pump will
continue to deliver pressure to the garment whilst
this warning is active.
After 10 minutes the warning changes to an alarm
state if the pump continues to detect that the
garment has been taken off the limb while the pump
is in the run state. The audible alarm is deactivated
by default for this alarm, but can be activated in the
software menu.
Reapply or replace the garment. The warning or
alarm will be cleared after a few minutes if the
garment is reapplied to the limb or immediately if the
middle control button is pressed to put the pump into
standby."
The service "spanner" icon appears at the right side
of the screen to indicate that the pump requires
service attention. This will normally occur after the
preset service interval has expired.
The LED indicators on the front and underside of the
pump remain green.
Therapy can continue until service is available.
Call the service engineer.
27