Guidance and manufacturer's declaration - electromagnetic immunity
Immunity Test
IEC 60601-1-2
Test Level
Electrostatic
±2kV, ±4kV, ±8kV,
discharge
±15kV air
(ESD)
EN 61000-4-2
±8kV contact
Conducted
3V in 0,15 MHz to
disturbances
80 MHz
inducted by RF
fields
6V in ISM and
amateur radio
EN 61000-4-6
bands between
0,15 MHz and 80
MHz
80% AM at 1 kHz
Radiated RF
Home Healthcare
electromagnetic
environment
field
10 V/m
EN 61000-4-3
80 MHz to 2,7 GHz
80% AM at 1 kHz
Electrical fast
±1kV SIP/SOP
transient/burst
ports
EN 61000-4-4
±2kV AC port
100 kHz repetition
frequency
Power
30 A/m
frequency
Magnetic field
EN 61000-4-8
Surge
±0,5kV ±1kV; ±2 kV,
AC Mains, Line to
Ground
±0,5kV ±1kV, AC
IEC 61000-4-5
Mains, Line to Line
Compliance
Electromagnetic environment - guidance
Level
±2kV, ±4kV,
Floors should be wood, concrete or ceramic
±8kV, ±15kV
tile. If floors are covered with synthetic
air
material the relative humidity level should be
at least 30%
±8kV contact
3V in 0,15
MHz to 80
MHz
Portable and mobile RF communications
6V in ISM
equipment should be used no closer to any
and amateur
radio bands
part of the product, including cables, than
1.0m, if the transmitter's output power rating
between
0,15 MHz
exceeds 1W
and 80 MHz
transmitters, as determined by an
electromagnetic site survey, should be less
80% AM at 1
than the compliance level in each frequency
kHz
b
range
Home
Interference may occur in the vicinity of
Healthcare
equipment marked with this symbol:
environment
10 V/m
80 MHz to
2,7 GHz
80% AM at 1
kHz
±1kV SIP/
Mains power supply should be that of a
SOP ports
typical commercial or hospital environment.
±2kV AC port
100kHz
repetition
frequency
30 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
±0,5kV ±1kV;
Mains power quality should be that of
±2 kV, AC
a typical commercial or hospital
Mains, Line
environment.
to Ground
±0,5kV ±1kV,
AC Mains,
Line to Line
35
a
Field strengths from fixed RF