Troubleshooting; General; Alarm Conditions; Troubleshooting Table - arjo Flowtron ACS900 Instrucciones De Uso

Ocultar thumbs Ver también para Flowtron ACS900:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7. Troubleshooting

General

On detection of a fault condition, the pump provides a visual-only warning followed
by an audible and visual alarm if the fault is not cleared.
Warning Conditions
The visual-only warning is a fault icon on the LCD screen. The LED indicators will
remain green, but may begin flashing, and there is no audible alarm.
The warning can be cleared by either:
• Rectifying the fault on the system, or
• Pressing the Middle Control button to put the pump into Standby.

Alarm Conditions

If the fault is not cleared then the warning is replaced by an audible and visual
alarm, which consists of:
• A fault icon on the LCD screen.
• The LED indicators on the front and underside of the pump change to yellow.
• An audible alarm will sound.
• All the alarm conditions are low priority.
The alarm can be cleared by either:
• Rectifying the fault on the system, or
• Pressing the Middle Control button to put the pump into Standby.
NOTE
For simplicity, the following screens (except Battery Low, Hardware Fail and
High Temperature) show a fault detected on garment 1; similar fault messages
are displayed if the fault is detected on garment 2 (or both garments).
NOTE
If the trouble shooting procedures do not return the system to normal
performance, stop using the system immediately and call the service engineer.

Troubleshooting Table

The following table provides typical warning and alarm conditions shown on
the LCD screen.
For each condition, there is a description and the relevant corrective action
required.
25

Publicidad

Tabla de contenido
loading

Tabla de contenido