9. Technical Specification

Pump Specification

PUMP
Part numbers
Supply Voltage (V)
Supply Frequency (Hz)
Power Input
Size
Weight
a. Configuration: STD is the Standard pump which has a standard length tubeset (2.1 m / 7 ft).
OR is the Operating Room pump which has a longer tubeset (4 m / 13 ft long).
Case Material
Mains Power Plug
Fuse Rating
Degree of protection
against electric
shock
Degree of protection
against liquid
ingress
Mode of operation
Pressure Range
ENVIRONMENTAL INFORMATION
Condition
Operating
Storage and Transport
(Long Term)
Storage and Transport
(Short Term)
NOTE
If the pump is stored in conditions outside of the "Operating" ranges, it should be allowed time to stabilise at
normal operating conditions before use.
H230 x W226 x D196 mm (9.1 x 8.9 x 7.5 in)
Flame Retardant ABS Plastic
5A to BS1362 (UK ONLY)
Class II, Double Insulated
Type BF
IPX3 - Protected against spraying water
Continuous
Foot Garment:
Uniform Calf and Calf & Thigh Garments:
Sequential (Tri Pulse) Calf and Calf & Thigh Garments:
Temperature Range
+10 °C to +40 °C
(+50 °F to +104 °F)
+10 °C to +40 °C
(+50 °F to +104 °F)
-20 °C to +50 °C
(-4 °F to +122 °F)
526000-XX
526000-17/18 (KSA)
526000-P-01/02 (America)
526000-P-31/32 (Russia)
100 - 230 V
230 V (KSA)
50 - 60 Hz
60 Hz (KSA)
10 - 40 VA
3.8 kg (8.5 lb)
Range:
Factory Default:
Relative Humidity
30% to 75%
(non-condensing)
20% to 95%
(non-condensing)
20% to 95%
33
130 ± 10 mmHg
35 - 65 ± 5 mmHg
40 ± 5 mmHg
45 ± 5 mmHg
Atmospheric Pressure
700 hPa to 1060 hPa
700 hPa to 1060 hPa
500 hPa to 1060 hPa