XII. PROBLEMS AND SOLUTIONS
The device does not work.
Check if the power cord from the adapter is plugged into the power network and all cable connectors
are correctly coupled. Use only the original adapter type FSP060-DAAN3.
Damaged power cord.
Do not use the device! Switch it off and replace the cable with new one or take the device to the
service facility.
Unpleasant or irritating thermal sensation by the patient.
Increase the distance of the device to the patient's teeth to 10-15 cm. If the thermal sensation is still
unpleasant at this distance switch to low intensity- "Lo" (Low) and shorten the distance to the teeth.
The display is blinking and no light is emitted.
The device is in PAUSE mode. To resume operation, press either (+) or (–) TIMER button.
The Bleaching Head cannot be fixed.
Tighten the fixing ball until the desired head position is reached (see Chapter V., point 5).
The display only shows "Oh".
The device has protection against overheating, and is activated if the temperature in the emitting
head rises. Then the light stops, the display shows "Oh" - Overheating - and the fans work for 1
minute. After cooling, the device returns to normal operation. Overheating can only occur in an
emergency or when the cooling openings are blocked. Therefore, the cooling openings in the lower
part of the radiating head must always be free for air circulation.
For all other questions regarding installation and operation of the BLANCONE
manufacturer or your local dealer.
16
Dental LED bleaching activation unit - MOBILE
TOUCH
STICK
HOME
DAY
CARE
ARCUS
, contact the
+
®
ULTRA
DUETTO
HOME
NIGHT
DUETTO
ACTIVE
CLICK
ARCUS
LIPS
ARCUS
DUETTO
CARE

Publicidad

Tabla de contenido
loading

Tabla de contenido