In the menu Trigger duration the pulse length for can be set to align with your medical
imaging recorder.
9. Endoscope Buttons Configuration
The Administrator and the Service Technician can set up
the endoscope button configuration, also when the endoscope
is not connected. All other users can view the current button
configuration, but not change it. The available functions depend
on the type of the selected endoscope. To view the current
configuration of a compatible endoscope or to reconfigure the
buttons, press the Settings tab in the tool bar, press Endoscope
buttons and select an endoscope type 1 . An overview will appear.
9.1. Configure the endoscope buttons
•
Press the Settings tab, then press Endoscope buttons.
•
Press the required endoscope category 2 and select an endoscope.
•
The screen shows an overview of the endoscope buttons with the available functions.
•
Press the required endoscope button 3 , then select long or short button press 4 .
•
Press the action to be performed when the selected button is pressed 5 .
•
Repeat for each button. The overview shows the selected functions next to the buttons.
2
Note: Each endoscope type comes with a default button configuration.
Note: The changes made will be saved and apply for all visualization devices of the same type.
Note: For some visualization devices e.g. ARC can only be assigned to be activated by
a "long press".
10. Operating the Displaying Unit
10.1. Preparation and inspection of the displaying unit before use
Before using the displaying unit for a patient procedure, follow the relevant steps below and
any other necessary steps according to your organization's procedures and requirements for
preparation and inspection of this type of device.