Guidance and manufacturer's declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Immunity test
Conducted Radio
Frequency
IEC 61000-4-6
Radiated Radio
Frequency
IEC 61000-4-3
Note 1: At 80 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast,
and TV broadcast, cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
IEC 60601-1-2
test level
3 V RMS
0,15 MHz –
80 MHz
6 V RMS
in ISM bands
80 % AM at 1 kHz
3 V/m 80 MHz to
2.7 GHz
80 % AM at 1 kHz
Compliance
Electromagnetic
Level
Environment Guidance
3 V RMS
Portable and mobile
0,15 MHz –
RF communications
80 MHz
equipment should be used
6 V RMS
no closer to any part of
in ISM bands
the system, including
80 % AM
its cables, than the
at 1 kHz
recommended separation
distance calculated from
3 V/m
the equation applicable
80-2700 MHz
to the frequency of
80 % AM
the transmitter.
at 1 kHz
Recommended
separation distance
d = 1.17√P
d = 1.17√P 80 MHz
to 800 MHz
d = 2.33√P 800 MHz
to 2.7 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey:
a) Should be less than the
compliance level in each
frequency range.
b) Interference may
occur in the vicinity of
equipment marked with
the following symbol.
45