Figure 4 Steep Needle Profiling with a curved array
To use a curved transducer in Needle Profiling mode
1 In 2D imaging, tap
Needle Profiling
Needle
Profiling
2 Ensure that the target is within the enhancement area (that is, within the outline) and that the
needle, when inserted, will approach the angle guide (dashed line) perpendicularly.
Tap
Left
or
Right
if you need to flip the image horizontally.
If this control is hidden, tapping
Tap
Shallow
,
Medium
The needle angle setting determines the location of the angle guide (dashed angled line).
Use the following table to determine the needle angle setting to use. Needle angles are
measured in relation to the surface of the transducer.
Table 4: Shallow, Medium, and Steep setting angles
Range for
Transducer
Shallow setting
C35xp
Less than 30°
C60xp
Less than 40°
.
More Controls
, or
Steep
to select a needle angle.
Range for
Medium setting
30°-40°
40°-50°
displays it.
Range for Steep
setting
Greater than 40°
Greater than 50°
13