8.3.3 Setpoint controller functions
Controller output
Limit
Hysteresis
Setpoint input
Proportional band X
p
Reset time T
n
Derivative action time T
Interpulse period
Minimum ON time T
min
Maximum frequency
Basic load
Maximum dosing rate
Control direction
Controller stop
8.4 Measuring ranges
CI
2
mg/l
0.00 - 0.50
0.00 - 0.50
0.00 - 1.00
0.00 - 1.00
0.00 - 2.00
0.00 - 2.00
0.00 - 5.00
0.00 - 5.00
0.00 - 10.00
0.00 - 10.00
0.00 - 20.00
The measuring ranges are also freely
Note
adjustable (within the above-mentioned
range limits).
Limit monitor, interpulse controller (P, PI, PID), pulse frequency controller (P,
PI, PID), continuous controller (P, PI, PID)
0 to 100 % of measuring range, adjustable in the unit of the measured value
(only with limit monitor)
0 to 50 % of full-scale value, adjustable in the unit of the measured value
(only with limit monitor)
0 to 100 % of measuring range, adjustable in the unit of the measured value
0.1 to 3000.0 %
1 to 3000 seconds, resolution 1 second (only with PI, PID controllers)
0 to 1000 seconds, resolution one second (only with PID controller)
v
1 to 100 seconds (only with interpulse controller)
0.1 to 10.0 seconds (only with interpulse controller)
1 to 180 pulses per minute (only with pulse frequency controller)
0 to 50 % of dosing rate
Value (basic load + 1) to 100 % of dosing rate
Adjustable to upward or downward control
Adjustable to NC or NO
CIO
O
2
3
mg/l
mg/l
0.00 - 0.50
0.00 - 1.00
0.00 - 2.00
0.00 - 5.00
H
O
pH
2
2
mg/l
pH
0-100
0.00 - 14.00
0-500
2.00 - 12.00
0-1000
5.00 - 9.00
0-2000
Pt100
C°
-5 to +120
13