11. Place the Charger above the Stimulator.
12. Slowly move the Charger down, towards the Stimulator.
13. Stop moving the Charger when it stops beeping.
14. Make a mark on your skin at the bottom edge of the
Charger (Figure 9).
15. Place the Charger below the Stimulator.
16. Slowly move the Charger up, towards the Stimulator.
17. Stop moving the Charger when it stops beeping.
18. Make a mark on your skin at the top edge of the Charger
(Figure 10).
The marks on your skin should create a square or
crosshair. The center of this crosshair indicates the center
of the Stimulator.
19. Now that you have located the center of the Stimulator,
properly align the Charger over the Stimulator (Figure 11).