Notes On The Safety Modes - Ottobock Genium X3 3B5-2 Instrucciones De Uso

Ocultar thumbs Ver también para Genium X3 3B5-2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 193
CAUTION
Improper mode switching
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Ensure that you stand securely during all switching processes.
Verify the changed damping characteristics after switching and observe the feedback from
the acoustic signal emitter.
Switch back to basic mode once the activities in MyMode have been completed.
Take the weight off the product and correct the switching, if required.
CAUTION
Improper use of the stance function
Falling due to unexpected product behaviour because of changed damping behaviour.
Make sure that you are standing safely when using the stance function and check the lock of
the knee joint before placing your full weight on the prosthesis.
Make sure you have been instructed in the proper use of the stance function by a prosthetist
or therapist. Information on the stance function see Page 68.

4.8 Notes on the safety modes

CAUTION
Using the product in safety mode
Falling due to unexpected product behaviour because of changed damping behaviour.
The warnings/error signals (see Page 87) have to be observed.
CAUTION
Safety mode cannot be activated due to malfunction caused by water penetration or
mechanical damage
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Do not continue using the defective product.
Contact your prosthetist immediately.
CAUTION
Safety mode cannot be deactivated
Falling due to unexpected product behaviour because of changed damping behaviour.
If you cannot deactivate safety mode by recharging the battery, a permanent malfunction has
occurred.
Do not continue using the defective product.
The product must be inspected by an authorised Ottobock Service Centre. The prosthetist
is your contact person.
CAUTION
Safety signal occurs (ongoing vibration)
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Observe the warnings/error signals (see Page 87).
Do not continue using the product after the safety signal has been emitted.
The product must be inspected by an authorised Ottobock Service Centre. In this case, the
contact person is your prosthetist.
60 | Ottobock

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Genium x3 3b5-2-st

Tabla de contenido