Overview Of Adjustment Parameters In Mymodes - Ottobock Genium X3 3B5-2 Instrucciones De Uso

Ocultar thumbs Ver también para Genium X3 3B5-2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 193
The following parameters can be modified:
Parameter
Resistance
Angle
Stance function
Sitting function
Stair Function
Pitch
Volume
In addition to the parameters, the remote control (optional accessory) can also be used to activate
deep sleep mode. Further information on deep sleep mode (see Page 80).

8.3.3 Overview of adjustment parameters in MyModes

INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
The parameters in the MyModes describe the static behaviour of the prosthesis for a specific
motion pattern such as cross-country skiing. Damping behaviour is not automatically controlled
and adjusted in MyModes.
Adjustment
App/remote
software
control
range
adjustment
range
120 – 180
+/- 10
55° – 70°
+/- 3°
Deactivated
0 - deactivated
Activated
1 - activated
Deactivated
0 - deactivated
Activated
1 - activated
Deactivated
0 - deactivated
Activated
1 - activated
1000 Hz –
1000 Hz –
4000 Hz
4000 Hz
0 – 4
0 – 4
Meaning
Resistance against flexion motion e.g.
when walking down stairs or when sit­
ting down
Maximum flexion angle in the swing
phase
Activation/deactivation of the stance
function. For switching with the Cock­
pit app or remote control, this function
has to be activated by the prosthetist.
Further information (see Page 68
).
Activation/deactivation of the sitting
function. For switching with the Cock­
pit app or remote control, this function
has to be activated by the prosthetist.
Further information (see Page 70).
Activating/deactivating the stair and
obstacle function. For switching with
the Cockpit app or remote control,
this function has to be activated by the
prosthetist. Further information (see
Page 70).
Pitch of beep signal for confirmation
tones
Volume of beep signal for confirma­
tion tones
Ottobock | 77

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Genium x3 3b5-2-st

Tabla de contenido