Descargar Imprimir esta página

Braun Aesculap Acculan 3Ti Instrucciones De Manejo/Descripción Técnica página 6

Cabezales
Ocultar thumbs Ver también para Aesculap Acculan 3Ti:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 76
1.
General information
1.1
Intended use
Task/Function
The Acculan 3Ti small drill GA671 or drill and reamer
GA672, combined with the relevant attachment and
tool, is used for processing hard tissue, cartilage and
the like, as well as bone replacement material, to screw
bone pins in and out and to set transfixion wires.
Application Environment
The product is used in operating rooms in sterile zones
outside of the explosion risk zone (such as areas with
pure oxygen or anesthesia gases).
1.2
Main functions and design charac-
teristics
Speed
GB628R
GB654R
GB655R
GB656R
GB658R
GB645R
GB176R
GB630R
GB634R
GB635R
GB636R
GB637R
GB638R
GB643R
Oscilla-
GB660R
tion fre-
quency
Rotational
Right and left rotation, oscillation
direction
4
-1
0 min
to max.
-1
250 min
-1
0 min
to max.
-1
900 min
-1
0 min
to max.
-1
1 000 min
-1
0 min
to max.
-1
17 000 min
Operating
Operation with non-periodic load and
mode
speed changes (type S9 pursuant to
IEC EN 60034-1)
Drilling (clockwise/counterclockwise
rotation):
60 second application, 60 second
pause
20 repetitions
30 min cooling time
Max. Temperature 48 °C
Reaming (clockwise/counterclockwise
rotation):
30 second application, 30 second
pause
8 repetitions
30 min cooling time
Max. Temperature 48 °C
Drilling (oscillation):
60 second application, 60 second
pause
4 repetitions
30 min cooling time
Max. Temperature 48 °C
Saw mode with GB660R:
30 second application, 60 second
pause
3 repetitions
30 min cooling time
Max. Temperature 48 °C

Publicidad

loading