12.7.2 Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfecting machine without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate
rinse
IV
Thermal disinfec-
tion
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, microbiological, at least of drinking water quality)
*Recommended: BBraun Helimatic Cleaner alcaline
►
For Kirschner wire attachment GB643R: set the
adjustment sleeve 4.2 to the greatest Kirschner
wire diameter.
►
Insert the product in its proper position in the
ECCOS fixation.
►
Then connect the interior rinse device to the ECCOS
fixation and connect to the rinse connector of the
cleaning/disinfection device/rinse cart.
►
After automatic cleaning/disinfection, check visible
surfaces for residues and repeat the cleaning/disin-
fection process as needed.
D
t
[°C/°F]
[min]
<25/77
3
55/131
10
> 10/50
1
90/194
5
-
-
Water
Chemistry/Note
quality
D–W
-
■
FD–W
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
working solution 0.5 %
– pH ~ 11*
FD–W
-
FD–W
-
-
mind. 10 min at max. 120 °C
en
19