14.2 Malfunctions of the Kirschner wire attachment
Malfunction
Kirschner wire cannot
be inserted into
Kirschner wire attach-
ment
Kirschner wire not
turning
14.3 Malfunctions of the sagittal saw attachment
Malfunction
Saw blade cannot be
uncoupled
15. Technical Service
Danger to life of patients and
users if the product malfunc-
tions and/or protective mea-
WARNING
sures fail or are not used!
For service and repairs, please contact your
national B. Braun/Aesculapp agency.
Modifications carried out on medical technical equip-
ment may result in loss of guarantee/warranty rights
and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
22
Detection
Kirschner wire diameter
not matching the setting
on the adjustment scale
Kirschner wire diameter
not matching the setting
on the adjustment scale
Clamping lever open
Cause
Release button not
pressed through prop-
erly
Do not perform any servic-
ing or maintenance work
under any circumstances
while the product is being
used on a patient.
Do not modify the product.
Cause
Incorrect adjustment of
Kirschner wire attach-
ment
Incorrect adjustment of
Kirschner wire attach-
ment
Clamping lever open
Detection
Locking pin not fully
releasing the saw blade
Other service addresses can be obtained from the
address indicated above.
Remedy
Set the adjustment sleeve to the
correct Kirschner wire diameter
Set the adjustment sleeve to the
correct Kirschner wire diameter
Close the clamping lever
Remedy
Fully press down the release
button, see Chapter 9.2.1