RECOMMENDED
INSERTION
PROCEDURE
SUMMARY
3
3. Irrigate.
6
6. Secure by tightening compression
cap. Verify waveform, reposition if
necessary.
4
After drilling hole in skull and penetrating the dura,
1
1. Screw in bolt.
4
4. Zero the transducer, then place
fingers at 5 cm and insert catheter
until fingers touch top of bolt.
7
7a
7. Hold catheter from above, pull
strain relief down and attach
to bolt.
7a. When the red depth indicator is
visible above the strain relief,
catheter tip (transducer) position
must be checked to ensure desired
depth is achieved.
2
2. Clear path with stylet.
5
5. Pull back slightly.
8
PREAMP
BOLT
CONNECTOR
8. To protect fiber optics, position
as shown and secure preamp
connector with clip.
CLIP