INTERNATIONAL MARKINGS
CATHETER EXTENSION CABLE
CONNECTIONS
Symbol also signifies:
TYPE BF EQUIPMENT: Protected against elec-
tric shock, having an F-type isolated (floating)
applied part.
Sterilized using ethylene oxide.
Date of expiration.
Do not reuse, single use device.
Manufacturer.
Do not resterilize.
Transducer size
Transducer type
Frequency response (system)
Model MPM-1/SPM-1
Model V420/M420/420XP
Model 420
Measurement range (system)
Zero drift (system)
First 24 hours (maximum)
5 days (typical)
Temperature coefficient
Linearity and hysteresis (system)
Pressure range:
-10 to 50 mmHg
51 to 125 mmHg
Reference pressure
Overpressure
Camino and Integra NeuroSciences are registered trademarks at Integra LifeSciences Corporation in the
United States and/or other countries. The Integra wave logo is a trademark of Integra LifeSciences
Corporation © 2010 Integra LifeSciences Corporation. All Rights Reserved.
MANUFACTURING INFORMATION
Integra NeuroSciences
5955 Pacific Center Blvd.
San Diego, CA 92121 • USA
(609) 275-0500 Facsimile (609) 275-5363
Customer Service (USA Only) - (800) 654-2873
having an F-type isolated (floating) applied part;
intended for direct cardiac application.
SPECIFICATIONS
4F (1.35mm Diameter)
Fiber optic
100Hz (-3dB)
120Hz (-3dB)
33Hz (-3dB)
-10 to 125 mmHg
0 ± 2 mmHg
less than ± 1 mmHg per day
Max of 3 mmHg over temperature range
of 22°C to 38°C (70°F-100°F)
± 2 mmHg or better
± 6% of reading or better
Atmosphere
-700 to 1250 mmHg
TYPE CF EQUIPMENT: Provided with a
degree of protection against electric
shock higher than Type BF equipment,
Refer to Directions for Use.
Lot Number.
Caution: Federal (USA) law
restricts this device to sale by
or on the order of a physician.
Authorized representative in the
European Community.
Integra NeuroSciences Ltd.,
Newbury Road • Andover • Hampshire • SP10 4DR
United Kingdom
Tel: +44 (0) 1264 345 700 Fax: +44 (0) 1264 332 113
5