3. Press In Scan to specify whether to include (On) or exclude (Off) the totalizer channel from the scan list.
4. Press Settings > Edge to specify if the totalizer channel should count on the rising edge (Pos) or falling edge
(Neg) if the input signal, and then press ReadReset to specify if the totalizer count should be automatically reset
(On) or not (Off) when read.
5. Press Read to read the totalizer count. The count is read once each time you press Read; the count does not
update automatically on the display.
6. Press Clear Count to reset the totalizer count to "0".
7. Press the Start/Stop softkey to toggle between Start to start the counting on the current totalizer channel, or
Stop to stop the counting on the current totalizer channel.
Keysight DAQ970A/DAQ973A User's Guide
3 Features and Functions