Descargar Imprimir esta página

Keysight DAQ970A Guia Del Usuario página 214

Publicidad

Idiomas disponibles

Idiomas disponibles

4     Measurement Tutorials
System Cabling and Connections
This section describes methods to reduce measurement errors that can be introduced by your system cabling. Many
system cabling errors can be reduced or eliminated by selecting the proper cable and grounding scheme for your
system.
Cable specifications
Grounding techniques
Shielding techniques
Separation of high-level and low-level signals
Source of system cabling errors
Cable specifications
A wide variety of general-purpose and custom cables are available. The following factors influence the type of cable
that you choose.
– Signal requirement - such as voltage, frequency, accuracy, and measurement speed.
– Interconnection requirements - such as wire sizes, cable lengths, and cable routing.
– Maintenance requirements - such as intermediate connectors, cable terminations, strain reliefs, cable
lengths, and cable routing.
Cables are specified in a variety of ways. Be sure to check the following specifications for the cable type you intend
to use:
– Nominal impedance (insulation resistance) - Varies with the frequency of the input signal. Check for HI-to-
LO, channel-to-channel, and HI-or LO-to-shield. High frequency RF applications have exact requirements
for cable impedance.
– Dielectric withstand voltage - Must be high enough for your application.
214
To prevent electrical shock or equipment damage, insulate all channels to the highest potential in the
system. It is recommended that you use a wire with at least 600 V rated insulation.
Keysight DAQ970A/DAQ973A User's Guide

Publicidad

loading

Este manual también es adecuado para:

Daq973a