3 Features and Functions
DAQM909A - 4 Channel 24-Bit Digitizer Module
The DAQM909A can be configured to digitize DC Voltage (DCV) and Strain. The following section describes how you
may configure the channels on the DAQM909A.
DC Voltage (DCV)
This section describes how to configure the channels on DAQM909A for DC voltage measurement from the front
panel.
1. Connect the voltage source to the module's screw terminals.
2. Press [Channel] > Measure, and select DCV from the selection menu
3. Press the Volt Range softkey to specify the DC voltage measurement range. You may choose between 300mV, 1
V (default value), 3 V,or 18V.
142
Keysight DAQ970A/DAQ973A User's Guide