• Make certain the unit is electrically grounded by connecting only to a grounded electrical service receptacle conforming to the
applicable national and local electrical codes.
• Transport 2 devices are designed to comply with electromagnetic safety standards. However, Transport 2 devices generate, use,
and can radiate radio frequency energy and, if not installed and used in accordance with instructions for use, may cause harmful
interference to other devices in the vicinity. Harmful interference to other devices can be determined by turning this equipment on
and off. Try to correct the interference using one or more of the following:
» Reorient or relocate the receiving device
» Increase the separation between the equipment
» Connect the equipment to an outlet on a different circuit from that to which the other device(s) are connected
» Consult your authorized DJO dealer for help
• Do not use Transport 2 Combo device in conjunction with electronic monitoring equipment (such as ECG monitors and ECG alarms).
Electronic monitoring equipment may not operate properly when electrical stimulation is in use.
• DO NOT operate Transport 2 in an environment where other devices are being used that intentionally radiate electromagnetic
energy in an unshielded manner. For example
» DO NOT operate the Intelect® Transport 2 within the vicinity or environment of an ultrasonic diathermy system.
» DO NOT operate the Intelect® Transport 2 within the vicinity or environment of any microware and RF shortwave
diathermy system.
• The energy from above systems can be transferred to patients with an implanted neurostimulation device causing tissue damage
and resulting in severe injury or death even if the implanted neurostimulation system is turned "off. "
• Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the Intelect Transport 2,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
• Simultaneous connection of a PATIENT to a high frequency surgical ME EQUIPMENT may result in burns at the site of the
STIMULATOR electrodes and possible damage to the STIMULATOR.
• DO NOT operate the Intelect Transport 2 unit when connected to any unit other than DJO devices or accessories specifically
described in this IFU as part of Transport 2 system or that have been specified as being compatible with the Transport 2. The use
of other companies' accessories, transducers or cables may result in increased emissions or decreased immunity of Transport
2 devices and resulting improper operations. DJO, LLC is not responsible for any consequence resulting from using products
manufactured by other companies.
• DO NOT apply stimulation transthoracically because the introduction of electrical current into the heart may cause cardiac
arrhythmia.
• DO NOT use contaminated, electrodes, lead wires, and gel which can lead to infection.
• DO NOT use electrode on multiple patients can lead to infection.
• Keep electrodes separated during treatment. Electrodes in contact with each other could result in improper stimulation or skin
burns.
• DO NOT apply electro stimulation treatment during bath, shower, sauna.
GENERAL WARNINGS AND PRECAUTIONS
WARNING
9