Program preselection
The PERIO keys (19) on the right side are prepro-
grammed with 3 basic settings.
When you briefly press one of these keys, the basic set-
tings are activated and the related information appears
on the display (12).
Factory-set values:
12
PERIO 1: Large rinsing media tank, 80 % power,
detection on, calculus signal tone, boost.
PERIO 2: Small rinsing media tank, 20 % power, detec-
tion on, calculus signal tone, boost.
19
PERIO 3: Small rinsing media tank, 50 % power, detec-
tion on, calculus signal tone, boost.
Reprogramming
You can assign new settings to the 3 PERIO keys:
Select e.g. the rinsing media supply type and power
value in %.
Press the desired PERIO key for more than 3 sec-
onds until the indicator bar next to the key flashes
briefly. The setting is saved.
Reset to factory settings
Press and hold the
ly switching the unit off and on again with the main
switch (1).
The unit is now reset to its factory settings.
Starting the PerioSonic ultrasonic handpiece
The light function and detection mode are automatically
activated when the handpiece is removed from its
holder. (Detection mode used only with PerioSonic
handpiece and SIROPERIO 4 PS or SIROPERIO PE8
tip).
The foot switch provides 3 different function modes:
For normal operation (9), abrasive power and cool-
ing/rinsing media, you must actuate both pedals si-
multaneously.
The boost function can be switched on with the left
pedal (10). The boost function can be activated only
with a power setting of at least 5 %.
In this mode, the preset power value is increased
by 20 %.
With a presetting of 80 % or higher, the maximum
'boost' value is always 100 %.
When you step on the right pedal (11), only rinsing
medium is discharged from the instrument tip, flush-
ing loose particles out of the treatment area.
The vibration of the instrument is deactivated in this
case.
Before switching to another function mode, briefly
release the foot switch.