5. Commissioning and operation
5.2 Electrical connection
Connection to built-in transmitter
For the electrical data (e.g. connection diagrams, output signal, measuring range, etc.)
please refer to the relevant operating instructions and/or data sheet for the built-in
head-mounted transmitter.
Junction between Ex d cable gland and connection head
■
Threads M20 x 1.5: tightening torques 20 Nm
Threads ½ NPT: tightening torques 20 Nm
→ Number of threads engaged according to the respective certification
Junction between cable and Ex d cable gland
■
Screw the male nut tightly into the adapter piece (use appropriate tools.)
During installation, take care to
Avoid distorting the cable sheath when tightening the male nut.
■
Avoid cutting too deep into the cable sheath.
■
Use suitable cable.
■
Be careful of the clamping area of the cable gland.
■
WIKA operating instructions, models TR10-L, TC10-L, TR10-0, TC10-0 (Ex d)