Table of Contents
1 Specifications
on page 3
2 General information
on page 4
3 Installation
on page 7
4 Startup
on page 16
5 User navigation
on page 16
Section 1 Specifications
Specifications are subject to change without notice.
Refer to the SC4500 Controller documentation for the SC4500 Controller specifications.
Specification
Dimensions
Weight
Material
Pollution degree
Overvoltage category
Protection class
Environmental conditions
Altitude
Operating temperature
Humidity
Sample
Calibration
Measurement range
Sensitivity
Repeatability
Reproducibility
Detection limit
Response time
Temperature compensation
Analog outputs (optional)
1
Dependent on controller configuration.
2
Refer to the module documentation for additional information.
Note: Install only one module in one of the available slots.
Details
Sensor with membrane—108 mm x ∅ 30 mm (4.25 x ∅ 1.18 inches)
Flow chamber—68 x ∅ 50 mm (2.68 x ∅ 1.97 inches)
Sensor with membrane—0.066 kg (0.15 lbs)
Flow chamber—0.74 kg (1.63 lbs)
Body and membrane holder: Noryl; body with optional immersion: stainless steel
316L; electrodes: gold cathode and silver anode; membrane: PFA; flow
chamber: stainless steel 316L
2
II
I, connected to protective earth
Indoor and outdoor use
2000 m (6562 ft) maximum
-20 to 60 °C (-4 to 140 °F)
0—95% relative humidity, non-condensing
Flow rate: 66 to 166 mL/min (4 to 10 L/h)
Temperature: -20 to 60 °C (-4 to 120 °F)
Pressure: atmospheric pressure
Zero—Electrically or with oxygen-free water
Slope—In air or against a laboratory measurement
0 to 2000 ppb (0 to 2 ppm); 0 to 45 °C (32 to 113 °F)
< 0.5 ppb
± 0.5 ppb or ± 5 % of measurement (the larger value)
± 0.5 ppb or ± 2 % of measurement (the larger value)
≤ 1 ppb
Change of 1 to 40 ppb: < 30 seconds
Automatic in the range 0 to 45 °C (32 to 113 °F)
1
Five 0–20 mA (or 4-20 mA) analog outputs on each analog output module
6 Operation
on page 16
7 Maintenance
on page 20
8 Troubleshooting
on page 23
9 Consumables and replacement parts
on page 27
2