Stryker Power-PRO IT Manual De Uso Y Mantenimiento página 103

Ocultar thumbs Ver también para Power-PRO IT:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

POWER-PRO™ IT (CONTINUED)
Recommended separations distances between portable and mobile RF
communication equipment and the Model 6516
The Model 6516 Power-PRO™ IT cot is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Model 6516 Power-PRO™ IT cot can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Model 6516 Power-PRO™ IT cot as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
www.stryker.com
EMC Information
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
D=(1.2)(√P)
0.12
0.38
1.2
3.8
12
6516-109-005 REV B
cot
Power-PRO™ IT
m
800 MHz to 2.5 GHz
D=(0.18)(√P)
0.018
0.57
0.18
0.57
1.8
Return To Table of Contents
English
D=(0.35)(√P)
0.035
0.11
0.35
1.1
3.5
1-97

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

6516

Tabla de contenido