Cot Retaining Post Replacement; Cot Retaining Post Screw Replacement - Stryker Power-PRO IT Manual De Uso Y Mantenimiento

Ocultar thumbs Ver también para Power-PRO IT:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

COT RETAINING POST REPLACEMENT

Tools Required:
• T30 Torx Driver
• 5/32" Hex Wrench
• Torque Wrench (in-lb)
Procedure:
1.
Raise the cot to the full upright position.
2.
Turn the cot onto the patient left side.
Note: Locate the arrow or groove on the bottom
bracket. The replacement retaining post bracket
will need to be assembled in the same orientation.
3.
Using a T30 Torx driver, remove the two socket
head cap screws (A) that secure the current cot retaining post to the base tube (B) (Figure 54). Discard the screws
and cot retaining post.
4.
Insert the button head cap screw (C) through the retaining post cap (D) and post tube (E), and then into the top
pin bracket (F) (Figure 54).
5.
Using a 5/32" hex wrench, tighten the button head cap screw (C) completely to secure the retaining post cap (D)
and post tube (E) to the top pin bracket (F) (Figure 54). Using a torque wrench, torque the screw to 100-140 in-lb.
6.
Assemble the cot retaining post across the base tube. Align the holes of the brackets and insert two socket head
cap screws (A) into the threaded holes of the bottom pin bracket (G) (Figure 54).
7.
Using a T30 Torx driver, tighten the two socket head cap screws completely.
8.
Verify proper operation of the unit before returning it to service.
Note: Adjustment of the rail clamp assembly may be required in order to compensate for any variation in cot retaining
post position depending on the ambulance cot manufacturer and model number.

COT RETAINING POST SCREW REPLACEMENT

Tools Required:
• T25 Torx Driver
• 5/32" Hex Wrench
• Torque Wrench (in-lb)
Procedure:
1.
Using a T25 Torx driver or 5/32" hex wrench, remove the button head cap screw that secures the retaining post
cap and post tube to the top pin bracket. Discard the screw.
2.
Using a 5/32" hex wrench, install and tighten the button head cap screw (0004-503-000) completely to secure the
retaining post cap and tube to the top portion of the retaining post assembly. Using a torque wrench, torque the
screw to 100-140 in-lb.
Note: If you cannot torque the screw to 100-140 in-lb, then you must replace the entire cot retaining post. See "Cot
Retaining Post Replacement".
3.
Verify proper operation of the unit before returning it to service.
www.stryker.com
Service Information
B
6516-109-005 REV B
C
D
E
F
G
A