7.4.3

Zero run on the Quick span machines

The procedure for the zero run is different to that of the
balance machines with the standard clamping means
in that the clamping sleeve is  tted on the shaft.
The shaft should be actuated outwards and the sleeve
placed on it up to the second groove from the left, after
which the shaft is actuated to the closed position.
7.4.4
Calibration run on Quick Span balance
machines
As with the zero run, the calibration procedure is done
with the clamping sleeve  tted on the shaft.
The shaft should be actuated outwards and the sleeve
placed on it up to the second groove from the left, after
which the shaft is actuated to the closed position.
1 689 979 917N