1) To switch off the Axon-Bus prosthethic system, press the charging receptacle button and hold
for more than one second (see Fig. 4, arrow).
2) Press both unlock buttons on the wrist unit (see Fig. 7).
INFORMATION: For safety reasons the Axon-Bus gripping component cannot be
removed if only one unlock button is pressed.
3) Pull off the Axon-Bus gripping component.
6.2 Attaching the Axon-Bus Gripping Component to the Socket
1) To switch off the Axon-Bus prosthethic system, press the charging receptacle button and hold
for more than one second (see Fig. 4, arrow).
2) Slide the Axon-Bus gripping component on to the socket until it locks in place audibly.
3) Verify proper attachment by pulling on the Axon-Bus gripping component.

7 Initial Operation

INFORMATION
Switching off the Axon-Bus prosthethic system during extended passive pauses (e.g. air or rail
travel, visit to a theatre or cinema, etc.) will make the battery last longer before requiring a
recharge. Only the entire Axon-Bus prosthethic system with all connected components can be
switched off. Individual Axon-Bus components cannot be switched off separately.
7.1 Switching On the Axon-Bus Prosthetic System
► Press the charging receptacle button and hold for one second (see Fig. 4, arrow).
→ The beeper emits two short signals.
7.2 Switching Off the Axon-Bus Prosthetic System
► Press the charging receptacle button and hold for more than one second (see Fig. 4, arrow).
→ The beeper emits one long signal.
7.3 Opening the Axon-Bus Gripping Component in an Emergency
This safety function allows the Axon-Bus gripping component to be opened regardless of the con­
trol signals present.
1) With the Axon-Bus prosthetic system switched on, press the charging receptacle button and
hold for approximately three seconds (see Fig. 4), until the Axon-Bus gripping component
begins to open.
→ A pulsating beep will sound as it opens.
2) Releasing the button immediately stops the process of opening the Axon-Bus gripping com­
ponent and turns the Axon-Bus prosthetic system off.
8 User Instructions

8.1 Storing the 8E500 Michelangelo Hand

CAUTION
Storing the 8E500 Michelangelo Hand in the closed position
Injury due to faulty control or malfunction of the 8E500 Michelangelo Hand because of damage
to the sensors and mechanism.
Store the 8E500 Michelangelo Hand in the open state (neutral position) only.
32 | Ottobock