Virtual SVV™ Instructions for Use - EN
Page 38
System function check
Before patient testing it is recommended that a quick check of the main system functions be performed.
Head tilt sensors
Tilt the virtual reality goggle to the right and to the left and check that a corresponding reading appears on
the software application device displays.
Virtual reality display
Initiate a measurement. Check that the luminous line is displayed in the virtual reality goggle.
Luminousline
Using the and buttons on the handheld remote, check that the luminous line rotates correspondingly
in the virtual reality goggle and in the application software.
Response confirmation
After a measurement has been initiated, press the OK button on the handheld remote to check that the
luminous line in the virtual reality goggle display extinguishes. Check that the measurement data is
entered in the results table and that the measurement data point appears in the graphical display.