Virtual SVV™ Instructions for Use – US
Page 16
To ensure a reliable wireless transmission it is important that a free line-of-sight is maintained between the
virtual reality goggle and base station. It is recommended that the base station is positioned in the vicinity
of the patient. To avoid electrical interference it should not be positioned close to the PC or monitor or any
other equipment likely to generate high frequency radiation.
The base station also functions as a charger for the handheld remote. For this purpose it should be
positioned so that the handheld remote can be inserted and held securely.
During charging in the base station the unit should be arranged so that the charge indicator LED can be
observed.
Access to the AC Power Adaptor must be ensured so that the AC Power cable can be pulled at any time.
Do not expose the device to direct sunlight or high heat supply.
3.6.5
Battery charging
During initial charging it is important that the handheld remote is kept continuously in the base station until
the yellow indicator LED extinguishes. Do not disconnect the base station from the AC power outlet
during this procedure.
Charging process
In order to charge the battery of the handheld remote insert it into the base station as shown in the figure
below. Charging is initiated automatically, and is indicated by the illuminated yellow LED on the handheld
remote. When charging is complete the indicator LED extinguishes.
Figure 3-14 Handheld remote in base station
During normal use the green ON LED blinks when the battery capacity has reached a critically low level
after 4 hours of continuous use. When this occurs, the handheld remote must be inserted into the base
station for recharging. Depending on battery capacity, recharging can take up to 5 hours.
NOTICE
If it is necessary to replace the batteries, please read and follow the safety instructions in Chapter 6.