Print Settings - Medizintechnik EasyOne-CS 2001 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 210
2030434-002 Rev. B
Relates to
Option
Current date
Current time
Alphanumeric
Yes, no
ID
Technician ID
Yes, no
Syringe volume
1.0l, 1.5l, ...7.0l
Height unit
m/cm, ft/inch
Weight unit
kg, lbs
Age, date of
Age, birth
birth
Contrast
Lang.
German, English,
others
Altitude above
0m, ...4000m
sea level
Op. mode
Diagnostic, Front-
line, NLHEP
Temp. unit
°C, °F
Rel. humidity
0...100%

8.4 Print Settings

Relates to
Options
Printer type
via PC
Result data
3 best values, best
values
Number of
3 best curves
Curves
best curve
Graph Types
FV&VT small, FV
large, VT large,
FV&VT large
Header 1-4
Optional entry
Default Setting
Description
Please enter the correct date at this point and
confirm with ENTER.
Please enter the correct time at this point and
confirm with ENTER.
No
If the ID you use also consists of letters, please set
to "Yes".
No
If you want the technician ID to be saved as well
and listed on the report, please choose "Yes".
3.0l
Choose the volume of your syringe if you wish to
conduct a calibration check.
m/cm
Choose how the unit will indicate height and alti-
tude.
North Am.:
ft/inch
Kg
Choose how the unit will indicate weight.
North Am.: lbs
Birth
If you use a database, consider entering the date
of birth so that the age can be calculated correctly
North Am.: age
at a later point.
Changes the display contrast.
English
Choose the desired language.
0m
Set the altitude above sea level of your location.
Diagnostic
see Chapter 8.1
°C
Determines how temperature is reported.
North Am.: °F
40%
Enter the average relative humidity at your loca-
tion.
Default Setting
Description
via PC
Set to 'via PC' by default if using CardioSoft
3 best values
You have the choice of printing out only the best
test or the 3 best tests on the report.
best curve
Choose if you want to print the 3 best curves of the
tests or only the best curve. It is only possible to
print the 3 best curves if the 3 best curves were
saved (see test settings "Save curve data").
FV&VT small
Choose what curves you wish to have on the re-
port.
Blank
You can enter the name and address of the institu-
tion or other information in 4 lines of 40 characters
each.