Electrical connection
7.2.3

Assembly of coaxial cable for pH / ORP electrodes

7.2.3.1 Single shielded connection cable

Pin assignment see
asymmetrical connection.
Fig. 12: Single shielded connection cable
1
Cable shield

7.2.3.2 Double-screened connection cable

Pin assignment see
balanced connection.
Fig. 13: Double-screened connection cable
1
inner cable shield
2
cable core (black)
The potential equalization line (brown) must only be connected if it has to be assumed
that the measurement device and measurement liquid are not on the same earthing
potential.
In this case, an additional electrical link to the measurement liquid is provided for potential
equalization and the sensor-side potential equalization line is to be applied there.
Fig. 14: Applying the grounding clamp
417102269 Rev. 2-08.2020
Ä Chapter 7.6.1 'pH/ORP (redox) measurement' on page 59
Ä Chapter 7.6.1 'pH/ORP (redox) measurement' on page 59
52
1
2
2
Cable core
1
3
2
4
3
Potential equalisation (brown)
4
4 outer shield