insTAlling The APPARATUs

This medical device is Intended for use by a qualified surgeon during endoscopic exploration.
No special training is required to install it. Kindly refer to the instructions in this manual.
5.1 insTAllATiOn
Place the apparatus on a stable surface and take the various accessories required for it to func-
tion;
If you are using an item of furniture, make sure it is well ventilated.
Connect the power cord to the mains socket of the apparatus [C3];
Check that the external filter is properly in place on the CO2 input socket [C6]. This is necessary
for the protection of the internal circuits and will help prolong the life of the insufflator.
5.2 COnneCTing TO A CYlinDeR OF MeDiCAl CO2
Maximum allowed operating pressure (60 bars)
Never begin any surgical operation without having a replacement cylinder of CO2
to hand. Only use medical-grade CO2 for coelioscopy, in compliance with European stan-
dards.
Place the cylinder of CO2 gas in a vertical position, so that it is secure with the up-
per part facing upwards, if possible on the same medical cart as the insufflator.
Never use the insufflator connected to a cylinder in a horizontal position or with the
upper part facing downwards.
Use the open-end spanner to tighten the connector attached to the cylinder outlet, and the
connector attached to the external filter on the input port on the insufflator [C6].
5.3 COnneCTing TO A MeDiCAl CO2 CenTRAl WAll POinT
Manually connect the low-pressure hose to the wall connector .
Minimum allowed operating pressure: 3 bars.
Only use with the CO2 gas central wall system, in compliance with European standards.