sPeCiAl FeATURes
This device can be connected to the CO2 system between 3 and up to 5 bars and to medical
CO2 cylinders. In the MENU, simply select "WALL POINT" or "CYLINDER".
If a CO2 cylinder is connected, the interface "Purge" button can be used to purge the high-pres-
sure hose before disconnecting it from the cylinder. To do so, remove the tubing hose from the
insufflator, close off the CO2 cylinder and press the "Purge" button.
A message is displayed on the screen when the insufflator detects a tubing hose obstruction.
This insufflator cannot function in two modes with differing maximum value settings for the pres-
sure setpoint.
The standard mode allows a maximum pressure setpoint value of 20 mm Hg. The bariatric
mode allows a maximum pressure setpoint value of 25 mm Hg.
7.11 CO2 CenTRAl WAll POinT AnD CO2 CYlinDeR
CAPACiTY On The sAMe DeViCe

7.12 PURge FUnCTiOn

7.13 hOse OBsTRUCTiOn MessAges

7.14 OPeRATing MODes