Technical Specifications Of The Ac440 Software - Interacoustics Affinity 2.0 Manual De Instrucciones

Ocultar thumbs Ver también para Affinity 2.0:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 417
3.2.4

Technical specifications of the AC440 software

Medical CE-mark:
Audiometer standards:
Transducers &
calibration:
Air conduction
DD45
TDH39
HDA300
DD450
HDA300
DD450
HDA280
E.A.R Tone
3A/5A
IP30
Bone
conduction
B71
B81
Free feld
High freqency
Effective
masking
Patient response switch:
Patient communication:
Monitor:
Stimuli:
Tone
Warble tone
Wave file
Masking
Narrow band
noise:
White noise:
Speech Noise.
Presentation
Intensity
Steps
Accuracy
Extended range
function
Frequency
Affinity2.0 – Instructions for Use - EN
The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II
of the Medical Device Directive 93/42/EEC. Approval of the quality system is made
by TÜV – identification no. 0123.
Tone: IEC60645-1/ANSI S3.6 Type 1
Speech: IEC60645-2/ANSI S3.6 Type A or A-E
Calibration information and instructions are located in the Service manual.
Check the accompanying Appendix for RETSPL levels for transducers
ISO 389-1 2017, ANSI S3.6-2018
ISO 389-1 2017, ANSI S3.6-2018
PTB report 1.61.4066893/13
ISO 389-8 2004, ANSI S3.6-2018
ISO 389-8 2006, ANSI S3.6-2010
ANSI S3.6-2018
PTB report 2004
ISO 389-2 1998, ANSI S3.6-2010
ISO 389-2 1998, ANSI S3.6-2018
Placemenet: Mastoid
ISO 389-3 2016, ANSI S3.6-2018
ISO 389-3 2016, ANSI S3.6-2018
ISO 389-7 2005, ANSI S3.6-2010
ISO 389-5 2004, ANSI S3.6-2010
ISO 389-4 1994, ANSI S3.6-2010
Hand held push button.
Talk Forward and Talk Back.
Output through external earphone or speaker.
Pure tone, Wable tone,NB, SN, WN, TEN noise
125-20000Hz separated in two ranges 125-8000Hz and 8000-20000Hz.
Resolution 1/2-1/24 octave.
1-10 Hz sine +/- 5% modulation
44100Hz sampling, 16 bits, 2 channels
Automatic selection of narrow band noise (or white noise) for tone presentation and
speech noise for speech presentation.
IEC 60645-1:2001, 5/12 Octave filter with the same centre frequency resolution as
pure Tone.
80-20000Hz measured with constant bandwidth
IEC 60645-2:1993 125-6000Hz falling 12dB/octave above 1KHz +/-5dB
Manual or Reverse. Single or multiple pulses. pulse time adjustable from 200mS-
5000mS in 50mS steps. Simultaneous or alternating.
Check the accompanying Appendix for maximum output levels
Available Intensity Steps is 1, 2 or 5dB
Sound pressure levels: ± 2 dB.
Vibration force levels: ± 5 dB.
If not activated, the Air Conduction output will be limited to 20 dB below maximum
output.
Range: 125Hz to 8kHz (Optional High Freqency: 8 kHz to 20 kHz)
Accuracy: Better than ± 1 %
Headband Static Force 4.5N ±0.5N
Headband Static Force 4.5N ±0.5N
Headband Static Force 8,8N ±0.5N
Headband Static Force 10N ±0.5N
Headband Static Force 8,8N ±0.5N
Headband Static Force 10N ±0.5N
Headband Static Force 5N ±0.5N
Headband Static Force 5.4N ±0.5N
Headband Static Force 5.4N ±0.5N