Descargar Imprimir esta página

Interacoustics Affinity 2.0 Manual De Instrucciones página 53

Ocultar thumbs Ver también para Affinity 2.0:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 417
Affinity2.0 – Instructions for Use - EN
Connect the CD player to the AUX1 input on the hardware.
Press START in the software and then tick the External sound check
box. The external sound will then play with the signal.
Use the slider to turn the sound level up and down.
Note that in Visible Speech Mapping you can select Live Voice and then
play an external sound. This means that you will have the external
sound alone with nothing (except from your own voice) interfering.
Current Protocol is listed in the lower left hand corner.
Upon performing a test the system automatically jumps to the next one
in the test flow. The checkmarks indicates that a curve has been
measured.
Test protocols can be created and adjusted in the REM440 setup.
Colour on each test button indicates the colour selected for each curve.
Start/Stop button initiates and ends the current test. Note that after
pressing START the text on the button will change to STOP.
The Graph shows measured REM curves. The X axis shows the
frequency and the Y axis shows the intensity of the test signal.
Gain/Response View allows for toggling between viewing the
curve as a gain or response curve. Note that this option is not
active for REIG.
Measurement Type is indicated above the graph, with a right/left
indication. In this example the REIG is displayed for the right ear.
Change the Input Level using the slider on the right hand side.
Scroll Graph Up/Down on the left hand side allows for scrolling
the graph up or down ensuring that the curve is always visible in
the middle of the screen.
Fitting Prescription and related details can be adjusted on the
right hand side of the screen. Select your preferred fitting
prescription in the upper dropdown list.
Choose between Berger, DSLv.5.0, Half Gain, NAL-NL1, NAL-
NL2, NAL-R, NAL-RP, POGO1, POGO2, Third Gain, or 'Custom'
if you have edited your target with the Edit feature.
Target shown will be calculated based on the selected fitting
prescription and the audiogram and can be shown as REIG
and/or REAR targets.
If no audiogram has been entered into
the audiogram screen, no targets will be displayed.
Note that fitting prescription settings (such as Age and Client
type) will differ depending on what fitting prescription is selected.