Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 8
Ending operation
Press the release key.
— Both LEDs flash green shortly and then red, accompanied by a
short signal tone.
— The pump is switched off.
Remove gas measuring device from the pump.
Observe the following during measuring mode with pump
The required waiting time when flushing the hose probe:
Before every measurement, flush the Dräger sampling hose or the
Dräger probes with the air sample to be measured.
— A flushing phase is necessary to eliminate or minimize all effects
associated with the use of a sampling hose or a probe, e.g. mem-
ory effect, dead volume.
— The duration of the flushing phase depends on factors such as
e.g. type and concentration of the gas or vapor to be measured,
material, length, diameter, and age of the sampling hose or probe.
Generally, when using a sampling hose (brand new, dry, clean), a
typical flushing time of approx. 3 seconds is required for each me-
ter. This flushing time applies in addition to the sensor response
time (see the Instructions for Use of the used gas measuring de-
vice).
Example:
— In the case of a sampling hose with a length of 10 m, the flushing
time is approx. 30 seconds and the sensor response time is in ad-
dition approx. 60 seconds. Therefore, the total time before read-
ing the gas measuring device is approx. 90 seconds.
— The flow rate alarm is delayed by 10 to 30 seconds depending on
the length of the hose.
Displays, Alarms, Remedies
LED / Signal Tone
Alarms
Remedies
Bat LED flashes red,
Battery alarm
Change batteries or
accompanied by a
charge rechargeable bat-
signal tone
teries.
Flow LED flashes red,
Flow alarm
Check sampling hose or
accompanied by a
probe for blockages.
continuous tone
Change filter on intake
port
Both LEDs flash red,
Device alarm
Have fault corrected by
accompanied by a
Dräger Safety Service.
signal tone
Replacing the Batteries
See "Commissioning" on page 4.
Changing the Filter
Unscrew the filter in a counterclockwise direction.
Unscrew the connecting bush from the filter.
Screw the connecting bush onto the new filter.
Screw the filter in a clockwise rotation onto the device.
Storing the Pump
Store the pump in a clean and dry place.
— It is recommended to store the pump separately from the gas
measuring device to ensure the flexible seal of the pump retains
its shape and sealing force.
In the case of long storage periods:
Remove batteries from the pump if the pump is not used for more
than 6 months. This avoids damage to the pump caused by leak-
ing batteries.
Disposing of the Device
EU-wide regulations for the disposal of electric and electronic
appliances which have been defined in the EU Directive
2002/96/EC and in national laws are effective from August
2005 and apply to this device.
Common household appliances can be disposed of using special
collecting and recycling facilities. However, as this device has not
been registered for household usage, it must not be disposed of
through these means. The device can be returned to your national
Dräger Safety Sales Organization for disposal. Please do not hesi-
tate to contact the above if you have any further questions on this
issue.

Technical Data

Ambient conditions
o
o
during operation
–20
C to 50
C
10 % r.h. to 90 % r.h.,
non-condensing
700 hPa to 1300 hPa
o
o
storage
–20
C to 50
C
10 % r.h. to 70 % r.h.,
non-condensing
700 hPa to 1300 hPa
Protection class
IP 64
Operation time
30 hours, with brand new alkaline batter-
ies, depending on the operating tempera-
ture and storage conditions
Power supply
3 alkaline batteries – T4 – (not rechargea-
ble!) Energizer No. E91, Energizer
No. EN91 (Industrial), Varta Type 4106
(power one), Varta Type 4006 or
3 NiMH rechargeable batteries – T3 –
(rechargeable) GP 180AAHC (1800 mAh)
o
max. 40
C ambient temperature
Flow rate
approx. 0.5 L/min
Maximum hose length
30 m
Alarms
optically by LEDs, acoustic signal tone
Dimensions
62 mm x 82 mm x 240 mm (W x H x D)
Weight
approx. 400 g
Approvals
(see "Notes on Approval" on page 46)
Order List
Designation and Description
Dräger Pump X-am 1/2/5000
Accessories / Spare Parts
Alkaline batteries "AA" batteries
Connection set filter
Complete cap
Sealing cap
Measuring probe 0.5 m
Measuring probe 1.5m
Plug-in telescopic probe
Telescopic probe 100 with accessories
Telescopic probe 150 stainless steel
Tester 90
Float probe with accessories
Viton hose
Hose (not suitable for H
S)
2
Case pump X-am 1/2/5000
Part No.
83 19 400
83 18 708
83 19 364
83 19 356
83 19 363
64 08 238
64 08 239
68 01 954
83 16 530
83 16 533
83 16 532
83 18 371
12 03 150
11 80 681
83 19 385
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pump x-am 2Pump x-am 5000

Tabla de contenido