Vercise™ DBS Remote Control 3 Directions for Use
From the Clinician Menu:
1. Select
Clear Link. The Clear Link to Stimulator? screen displays.
2. Select
Yes to clear the link to the Stimulator.
or
Select
No to cancel the action.
Impedances
You can use the Remote Control to check impedances.
From the Clinician Menu:
1. Select
Impedances. An impedance measurement is taken and the Impedances screen is
displayed.
When an impedance measurement is taken, impedances are assessed between a contact and the
case (monopolar), and between pairs of contacts (bipolar). A green square indicates that impedance is
within the acceptable range. A yellow dot with a question mark indicates that impedance is outside of
the acceptable range.
Vercise™ DBS Remote Control 3 Directions for Use
91172931-02 REV A 16 of 835