Boston Scientific Vercise DBS Instrucciones De Uso página 30

Ocultar thumbs Ver también para Vercise DBS:
Tabla de contenido

Publicidad

Vercise™ DBS Remote Control 3 Directions for Use
Charge Stimulator Now or Disable MRI Mode Screen
If MRI Mode has already been enabled and the Stimulator battery power falls below the recommended
value, the Remote Control will display a message instructing the patient to charge the Stimulator.
To charge the Stimulator without disabling MRI Mode:
1. Do not press
.
2. Instruct the patient to charge the Stimulator.
3. Check the Remote Control to confirm that the error message has cleared.
4. Navigate to the Home Screen on the Remote Control by pressing
Control and confirm that the MR Conditional Symbol
The patient can also disable the MRI Mode before charging the Stimulator:
1. Press
to disable the MRI Mode.
2. Instruct the patient to fully charge the Stimulator.
3. Check the Remote Control to confirm that the error message has cleared.
4. Enable MRI Mode by following instructions in the Enabling MRI Mode section in this manual.
Caution: Charger and Remote Control are MR Unsafe and must not be brought into the MRI scanner
room.
Vercise™ DBS Remote Control 3 Directions for Use
91172931-02 REV A 26 of 835
on the side panel of the Remote
is displayed on the Home Screen.