Vercise™ DBS Remote Control 3 Directions for Use
8. If MRI Mode is disabled, the MRI Mode Disabled confirmation screen is displayed.
Note: The Stimulator will retain the stimulation and program settings that were set before MRI Mode
was enabled. If the stimulation was ON before MRI Mode was enabled, then disabling MRI
Mode turns the stimulation back ON. If stimulation was OFF before MRI Mode was enabled,
then disabling MRI Mode keeps the stimulation OFF.
9. The Home Screen on the Remote Control will not display the MR Conditional Symbol once MRI Mode
is disabled.
MRI Mode Error Screens
The Remote Control performs system checks once "Enter MRI Mode" is selected from the Systems
Settings. It will display error screens if :
• The Stimulator battery is not fully charged.
• The Impedance check detects an anomaly.
• There is an error in the Stimulator.
Vercise™ DBS Remote Control 3 Directions for Use
91172931-02 REV A 24 of 835