Hemolysis: Presence of packed red blood
cells in concentrations up to 30 µL/mL has
no effect on results, within the precision of
the assay.
Lipemia: Presence of triglycerides in
concentrations up to 3000 mg/dL has no
effect on results, within the precision of the
assay.
Method Comparison: The assay was
compared to Coat-A-Count
patient samples. (Concentration range:
approximately 3.3 to 41 µg/dL. See
graph.) By linear regression:
(IML) = 0.95 (CAC) + 0.32 µg/dL
r = 0.988
Means:
14.6 µg/dL (IMMULITE)
15.0 µg/dL (CAC)
References
1) Foster L, Dunn R. Single-antibody technique
for radioimmunoassay of cortisol in unextracted
serum or plasma. Clin Chem 1974;20:365.
2) Farmer R, Pierce C. Plasma cortisol
determinations: radioimmunoassay and
competitive protein binding compared. Clin
Chem 1974; 20:411. 3) Ruder H, et al. A
radioimmunoassay for cortisol in plasma and
urine. J Clin Endo Metab 1972;35:219.
4) Rothfeld B, ed. Plasma Cortisol. In: Nuclear
medicine in vitro. 1974:120. 5) Murphy B, et al.
Clinical studies utilizing a new method for the
serial determination of plasma corticoids. J
Canad Med Assoc 1964;90:775. 6) Sparks R.
Measurement of serum 11-deoxycortisol and
cortisol after metyrapone. Ann Intern Med
1971;75:717. 7) Kowalski A, Paul W. A simple
extraction procedure for the determination of
free (unconjugated) cortisol in urine by
radioimmunoassay. Clin Chem 1979;25:1152.
8) Burtis CA, Ashwood ER, editors. Tietz
textbook of clinical chemistry. 2nd ed.
Philadelphia: W.B. Saunders, 1994. 9) Wilson
JD, Foster DW, editors. Williams textbook of
endocrinology. 7th ed. Philadelphia: W.B.
Saunders, 1985. 10) National Committee for
Clinical Laboratory Standards. Procedures for
the collection of diagnostic blood specimens by
venipuncture; approved standard. 4th ed.
NCCLS Document H3-A4, Wayne, PA: NCCLS,
1998. 11) Centers for Disease Control. Update:
Universal precautions for prevention of
transmission of human immunodeficiency virus,
hepatitis B virus and other bloodborne
pathogens in healthcare settings. MMWR,
1988;37:377–82, 387–8. 12) Clinical and
Laboratory Standards Institute (formerly
NCCLS). Protection of Laboratory Workers From
Occupationally Acquired Infections; Approved
Guideline - Third Edition. Wayne, PA: Clinical
and Laboratory Standards Institute; 2005.
IMMULITE/IMMULITE 1000 Cortisol (PILKCO-16, 2016-02-08)
NCCLS Document M29-A3. 13) Federal
Occupational Safety and Health Administration,
Bloodborne Pathogens Standard, 29 CFR
1910.1030.

Technical Assistance

In the United States, contact
Siemens Healthcare Diagnostics
Technical Services department.
Tel: 877.229.3711. Outside the United
®
Cortisol on 83
States, contact your National Distributor.
siemens.com/healthcare
The Quality System of Siemens Healthcare
Diagnostics Products Ltd. is certified to ISO
13485:2003.
Tables and Graphs
Precision (µg/dL)
1
Within-Run
3
4
5
Mean
SD
CV
1
2.7
0.24
8.8%
2
4.4
0.33
7.5%
3
11.3
0.88
7.8%
4
19.0
1.1
5.8%
5
29.1
1.8
6.2%
6
35.8
2.4
6.7%
2
Total
SD
CV
0.27
10%
0.37
8.4%
0.87
7.7%
1.2
6.3%
2.2
7.6%
2.6
7.3%
5