If any protruding parts of the Delivery Housing are present or any clip points extend beyond distal rim/edge,
carefully replace safety cap, DO NOT use this product, contact your local STERIS Endoscopy Product
Specialist or Customer Service Representative, and use a different device for procedure.
7.
Carefully apply lubricant liberally and circumferentially to the outside of the Delivery Housing.
8.
Set-up is now complete.
Caution:
o
Keep hands and fingers clear from the tip of the Delivery Housing while handling the endoscope.
Do Not Remove the Handle Stay until endoscope with loaded Padlock Clip
o
over defect and ready for clip deployment.
Device Insertion into patient:
1.
Intubate patient with the endoscope and the mounted Padlock Clip
standard endoscopic techniques for procedures utilizing an over the scope clip.
Note:
o
Use of an overtube, such as the Guardus
traverse areas within the upper gastrointestinal tract and minimize the risk of mucosal injury.
o
Placement of the Guardus
endoscope prior to the use of the Padlock Clip
o
Once the Guardus
can be safely inserted into the patient and directed to the desired location of treatment.
o
Physicians may also choose to dilate any stenosed area prior to passing the Padlock Clip
o
The outer diameter of the Padlock Clip
Clip Pro Select
732380 Rev. K
®
overtube should be performed with a separate maneuver and appropriately sized
For online set up video and complete instructions for use for the Guardus
following URL into your browser's search bar:
http://www.usendoscopy.com/products/guardus-overtube-esophageal.
®
overtube is in place, the endoscope with the mounted Padlock Clip
The inner diameter of the Guardus
defect closure system is 16mm.
Carefully apply lubricant liberally and circumferentially to the outside of the Delivery Housing prior to
insertion through the Guardus
®
defect closure system is 19mm.
®
defect closure system and direct to desired location using
®
overtube, may be warranted to help the Padlock Clip
defect closure system.
®
®
overtube is 16.7mm and the outer diameter of the Padlock Clip
®
overtube.
®
defect closure system is 16mm and the outer diameter of the Padlock
®
defect closure system is in position
®
delivery system safely
®
overtube click or paste the
®
defect closure system
®
delivery system if necessary.