Plugging or unplugging the product

WARNING
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
Always store the power cord to avoid the risk of entanglement, damage to the power cord, or potential shock
hazards. If the power cord is damaged, remove the product from service and contact the appropriate maintenance
personnel.
Always allow enough clearance between the head end of the product and the adjacent wall, so you can unplug the
power cord from the wall outlet during an emergency.
Always store the power cord before transporting the product.
To plug in the product, plug the power cord into a protective earthed outlet.
To unplug the product, grasp the mold near the outlet and pull the cord in a direction parallel to the floor (not at an
angle).

Charging the battery

WARNING
Always disconnect the power cord from the wall outlet if you detect overheating of the battery, control cables, or
pendants. Do not use the product again until it has been inspected, serviced, and confirmed to work as intended by
authorized maintenance personnel.
Always replace the battery after it surpasses its expected service life.
Do not open a dead battery.
Do not throw the battery into a fire.
Do not spill liquid onto the battery or submerge the battery in liquid.
CAUTION
Always plug the product into a wall outlet (regulated AC-power source) when not in use to maintain a sufficient
battery charge and to maximize product performance while operating on battery power.
Always replace batteries that have corrosion at the terminals, display cracking, have expanded or bulging sides, or
no longer can maintain a full charge.
Only use authorized batteries when replacing the batteries. Use of unauthorized batteries may lead to unpredictable
system performance.
SV2 is equipped with a battery backup system that charges when the product is plugged into a wall outlet. The battery
backup system allows the operator to use the product when the product is unplugged, during a power failure, or while
transporting a patient. Battery backup functionality activates when you unplug the product.
Always check battery backup function according to the preventive maintenance checklist (see
on page
1-44). Always replace the battery if it does not perform as intended during preventive maintenance.
www.stryker.com
Operation
KK-7500 REV 02
Preventive maintenance
1-17
English
EN