Preparing the product for cleaning

Cleaning and disinfecting are two separate processes. Clean before disinfecting to make sure that the cleaning agent is
effective.
To prepare the product for cleaning:
1.
Raise the litter to its highest height.
2.
Lock the siderail control panel and patient control pendant functions (see
3.
Unplug the power cord from the wall outlet.
4.
See
Storing the power cord on page
5.
See
Applying or releasing the brakes on page
6.
Remove the mattress.

Cleaning

WARNING
Do not clean, service, or perform maintenance while the product is in use.
Always power off and unplug the power cord before cleaning, servicing, or performing maintenance.
Always power off the product and unplug the power cord from the wall outlet when large spills occur near the circuit
boards, cables, and motors. Remove the patient from the product, clean up the fluid, and have service personnel
inspect the product. Fluids can cause unpredictable operation and decreased functionality of any electrical product.
Do not return the product to service until it is completely dry and has been thoroughly tested for safe operation.
Do not spray cleaners directly onto the battery, control boxes, actuators, cables, or other electric equipment.
Do not use abrasive powder, steel wool, or similar materials that may damage the product surface.
Do not use Virex
®
TB for product disinfecting.
Do not use acid-based chemicals or flammable chemicals, such as gasoline, diesel, or acetone for cleaning
purposes.
Do not directly spray or saturate the siderail control panel, patient control pendant, or nurse control pendant with
cleaners.
The cleaners and disinfectants must not be highly alkaline or acidic (pH value 6-8).
CAUTION
Do not steam clean, pressure wash, ultrasonically clean, or immerse any part of the product in water. Exposure to
water may damage the internal electric parts. These methods of cleaning are not recommended and may void this
product's warranty.
Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning.
Some cleaning products are corrosive in nature and may cause damage to the product if you use them improperly.
If you do not properly rinse and dry the product, you may leave a corrosive residue on the surface of the product
that could cause premature corrosion of critical components. Failure to follow these cleaning instructions may void
your warranty.
To clean product surfaces:
1.
Using a clean, soft, damp cloth, wipe product surfaces with a mild soap and water solution to remove foreign
material.
2.
Wipe product surfaces with a clean, dry cloth to remove any excess liquid or cleaning agent.
3.
Dry thoroughly.
www.stryker.com
Cleaning
1-18.
1-19.