Immunity test
Test level according to:
Electrostatic dis-
± 8 kV Contact
charges (ESD)
± 15 kV Air
(IEC61000-4-2)
Fast Transients
± 2 kV Input a.c. Power
(IEC61000-4-4)
Ports
± 1 kV Input Power Ports
Surge Immunity
Test
(IEC61000-4-5)
± 2 kV Signal Input /
output
Power-frequency
30 A/m
magnetic field
(IEC61000-4-8)
0% UT
0.5 cycles
A 0°, 45°, 90°, 135°, 180°,
225°, 270° et 315°
Voltage Dips
0% UT
and Interruptions
1 cycle
(IEC61000-4-11)
70% UT
25 cycles 50 Hz
30 cycles 60 Hz
Single phase: à 0°
Voltage Dips and
0 % UT;
Interruption
250 cycles 50 Hz
(IEC61000-4-11)
300 cycles 60 Hz
Electromagnetic emissions
Level of conformity
IEC60601
± 8 kV Contact
± 15 kV Air
± 2 kV Input a.c. Po-
wer Ports
± 1 kV Signal Input/
Output Ports*
± 1 kV Input Power
Ports
± 2 kV Signal Input / out-
put
30 A/m
0% UT
0.5 cycles
A 0°, 45°, 90°, 135°,
180°, 225°, 270° et 315°
0% UT
1 cycle
70% UT
pour 25 cycles à 50 Hz
Pour 30 cycles à 60 Hz
Single phase: à 0°
0 % UT;
250 cycles 50Hz
300 cycles 60 Hz
226
Electromagnetic environ-
ment / remarks
Home health care environ-
ment and an environment of
a professional health care
facility.
Home health care environ-
ment and an environment of
a professional health care
facility.
Home health care environ-
ment and an environment of
a professional health care
facility.
Home health care environ-
ment and an environment of
a professional health care
facility.
Home health care environ-
ment and an environment of
a professional health care
facility.
Home health care environ-
ment and an environment of
a professional health care
facility.