Plug the SpO
probe into USB interface, as Figure 4-3-2-4.
2
1) Connect the probe with the device.
2) Put the finger into the probe.
3) In measurement interface,you can directly read corresponding data from dis-
play screen.
Warning: Use the SpO
probes are forbidden.
Notes:
Testee's fingernail shall not too long.
Testee shall not use enamel or other makeup.
The finger too cold or too thin may affect the measurement; when measuring,
please insert the thicker finger fully into the probe (thumb or middle finger rec-
ommended).
In operation, the finger tested shall not tingle and the testee shall not be in
movement.
When inserting the finger, fingernail and the luminescent tube should be on the
same side
ENGLISH
Figure 4-3-2-3 Binding SpO
Figure 4-3-2-4 Placing SpO
probe specified by our company,and other SpO
2
probe
2
probe
2
18