 Excessive ambient light may affect the measuring result,including fluorescent
lamp, dual ruby light, infrared heater, direct sunlight and etc.
 There should not be such light barrier as rubber fabric on the way of light, or it
may result in incorrect measure results of SpO
 The SpO
probe and photoelectric receiving tube should be arranged in a way
2
with the testee's arteriole in a position there between.
 The SpO
sensor should not be used at a location or limb tied with arterial
2
canal or blood pressure cuff or receiving intravenous injection.
 Acute testee movement or electrosurgical interference can affect the accuracy.
 Data update period is less than 5 seconds, which is changeable according to
different individual pulse rate.
 Please read the measured value when the waveform on screen is equable and
steady-going, of which the value is optimal. And the waveform at the moment
is the standard one.
 The problem of overrating would display when the patient is suffering from
toxicosis which caused by carbon monoxide, the device is not recommended
to be used under this circumstance.
 The finger shall be put correctly (refer to the related Figure 4-3-2 in the manual)
or the measurement results may be incorrect.

4.4 Clinical Restrictions

SpO
2
1) As the measure is taken on the basis of arteriole pulse, substantial pulsating
blood flow of subject is required. For a subject with weak pulse due to shock,
low ambient/body temperature, major bleeding, or use of vascular contracting
drug, the SpO
ment will be more sensitive to interference.
2) For those with a substantial amount of staining dilution drug (such as methyl-
ene blue, indigo green and acid indigo blue), or carbon monoxide hemoglobin
(COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with
icterus problem, the SpO
3) The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may
also be a major factor blamed for serious error of SpO
4) As the SpO
value serves as a reference value for judgement of anemic anoxia
2
and toxic anoxia, some patients with serious anemia may also report good
SpO
measurement.
2
Warning: Please refer to corresponding medical literature for details about
clinical limits and tabu illness.
waveform (PLETH) will decrease. In this case, the measure-
2
determination by this monitor may be inaccurate.
2
ENGLISH
and pulse rate.
2
measure.
2